Four heating elements are located on top bottom and sides of chamber for enhanced temperature uniformity.
Built-in vent port removes undesirable contaminants and moisture; extends the life span of the element and unit.
The rear of the chamber incorporates a 0.38" (0.95 cm) diameter port for monitoring chamber temperatures with independent measuring devices.
Safety
Optional stainless steel shelf doubles load capacity (maximum temperature of 900°C).
Door safety switch protects operator by removing power to the heating elements upon opening the door.
Thermocouple break protection cuts power to the heating elements prevents a thermocouple failure runaway condition.
Models F6010 and F6018 are supplied with cord and plug set all others require conduit connection.
Applications
Ashing organic & inorganic samples.
Gravimetric analysis.
Operation
Four temperature controller models:
- A1: Digital single set point control.
A single display shows actual temperature or set point.
- B1: Digital single set point control with a single ramp to set point and a dwell.
A single display shows Actual temperature or set point.
The furnaces that use this control also utilize a mechanical Over Temperature Protection relay.
- C1: Digital programmable control with one stored program of 8 segments.
The furnaces that use this control also utilize a mechanical Over Temperature Protection relay.
- D1: Digital programmable control with 4 stored programs 16 segments per program.
The controller also includes RS232 communications.
The furnaces that use this control also utilize a mechanical Over Temperature Protection relay.
Chamber Dimensions W x D x H in (cm) 12.8 (33) x 10 (25) x 6.8 (18) Overall Dimensions W x D x H in (cm) 19.1 (49) x 20 (51) x 21 (53) Chamber cu. ft. (L) 0.5 (14)