Speeds: Choice of 3 options. Instrument has “Interleaved” speeds when manufactured. Standalone Interleaved: LV/RV (18 speeds) 8 LV speeds followed by 10 RV speeds Sequential: LV/RV (18 speeds) 8 LV speeds and 10 RV speeds arranged in sequential order from lowest 0.3 rpm to highest 100 rpm. Custom: 54 speeds, user selectable External (PC Control) 0.01 - 200 rpm 0.01 rpm increments from 0.01 to 0.99 rpm 0.1 rpm increments from 1.0 to 200 rpm
Temperature Sensing Range: -100C to 300C (-148F to 572F) Analog Torque Output: 0 - 1 Volt DC (0 - 100% Trque) Analog Temperature Output: 0 - 3.75 Volts DC (-100°C to 275°C) RS232 Compatible Serial Port for use with an attached printer or PC. Centronics Compatible Parallel Port for use with an attached printer. Viscosity Accuracy: ±1.0% o full scale range Viscosity Repeatability: ±0.2% o Full Scale Range
Temperature Accuracy: ±1°C | -100°C to 149°C ±2°C | 150°C to 300°C Operating Environment: 0°C to 40°C temperature range (32°F to 104°F) 20% -80%R..: non-condensing atmosphere