CEM Microwave Digestion System MDS-2100(USA)
상태양호
전원 220
연락처 011-354-6248
Specifications:
Moisture/Solids Range = 0.01% to 99.99% in liquids solids and slurries. 0.01% resolution.
Balance Capacity = 50 Grams 0.1mg Readability
Microwave Power = 0 to 100% of full power (1000 watts) in 1% increments.
Program Storage = 20 methods
Standard Software = Constant weight Fat Analysis (Fat Tester Fat Analyzer) Food Meat Wastewater Dilutions Syringe Weighing.
Data Entry = Alphanumeric Keypad with Menu-driven software
Display = High resolution backlit LCD graphics display
Printer Port = 25-pin parallel type
Computer Port = RS232 9-pin IBM AT compatible
External Balance Port = RS232 9-pin
Electrical Requirements = 108/125 VAC 60 Hz 15 Amps. 220/250 VAC 50 Hz 10 Amps.
Instrument Dimensions = 21.4 inches x 18.1 inches x 22.5 inches
Weight = 110 lbs
cem 9000 에서 응용함
Samples for trace metals analysis are prepared and placed into specialized Teflon lined microwave digestion vessels. The vessels are placed in the CEM MDS-2100 system for microwave digestion. The MDS allows a programmable temperature and/or pressure profile that can be monitored during the digestion procedure to insure complete digestion of samples. Twelve samples can be digested in a single run with temperature and pressure being monitored for assurance of complete digestion. All sample information temperature and pressure profiles are printed on hardcopy at the time of digestion for easy review and to complete data sets.